| Model Number |
LDC315T |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Quest medical, inc.Has limited information related to the patient's medical history and is unable to form an opinion as to the relevancy of the patient's history to the event reported.Quest medical, inc.Defers to the patient's physician regarding medical history.
|
| |
|
Event or Problem Description
|
|
The facility materials manager reported an issue encountered by the physician while using the lacrimal duct catheter.The report stated that the device would not remain inflated.The report stated that the catheter was removed and a second one was used to successfully complete the procedure.There were no patient complications reported as a result of the alleged issue.The catheter was returned to the manufacturer for evaluation.
|
| |
|
Additional Manufacturer Narrative
|
|
Because the complaint sample was a device included in a kit,, there were 2 possible lots of product it could have been.The other possible lot of product is lot 0523966n02 manufacture date 10/26/2016 exp 10/26/2019.Visual examination of the complaint sample found that it had been used.After decontamination, the device was inflated to 8atm.The device remained inflated during pressurization; however, once completed a leak was observed at the tip.The root cause is most likely inadequate adhesive to bond the balloon to the catheter body.The possible product lots were both manufactured prior to manufacturing process improvements implemented.
|
| |
|