| Catalog Number |
03L80-31 |
| Medical Device Problem Code |
Low Test Results (2458)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
01/16/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Lot/serial number was manufactured prior to udi compliance date; therefore, only a di is provided.Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs and a review of labeling.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.A malfunction was identified as the device failed to meet performance specifications or otherwise perform as intended at the customer site.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
|
| |
|
Event or Problem Description
|
|
The customer observed falsely depressed carbon dioxide results on the architect c4000 analyzer.The following data was provided: patient 8 initial 11, repeats 23 meq/l there was no known impact to patient management.
|
| |
|