| Catalog Number |
48335312 |
| Medical Device Problem Codes |
Mechanical Problem (1384); Migration or Expulsion of Device (1395)
|
| Health Effect - Clinical Codes |
Injury (2348); No Information (3190)
|
| Date of Event |
03/14/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that screw backed out, 4months post op.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection; device history review; complaint history review; risk assessment.Some deformities were found under the locking clip and on the top of the screw.Manufacturing review was performed and no anomalies were identified.The plausible root cause cannot be determined based on the information provided.
|
| |
|
Event or Problem Description
|
|
It was reported that screw backed out, 4 months post op.
|
| |
|
Search Alerts/Recalls
|