| Model Number |
4542 |
| Medical Device Problem Codes |
Pacing Problem (1439); Extracardiac stimulation (1463); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
Muscle Stimulation (1412)
|
| Date of Event |
02/16/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The product has been received for analysis.This report will be updated upon completion of analysis.
|
| |
|
Event or Problem Description
|
|
Boston scientific received information that this left ventricular (lv) lead exhibited pacing inhibition and was found to be dislodged one day post implant.The patient experienced stimulation, however, did not experience any symptoms as a result of the pacing inhibition.An invasive procedure was performed.The lead was explanted and replaced with another manufacturer's lead.No additional adverse patient effects were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Upon receipt at our post market quality assurance laboratory, a thorough evaluation of the lead was performed.Resistance and pressure tests were completed to assess lead electrical performance and insulation integrity.Measurements throughout these tests were within normal limits.Microscopic inspections of the terminal pin assembly, lead body, and electrode tip found no anomalies.Laboratory testing was unable to reproduce the reported clinical observations and detailed analysis did not reveal any abnormalities.Laboratory analysis did not identify any lead characteristics that would have caused or contributed to the reported clinical observations.
|
| |
|