Analysis was completed on the returned lead portion.A large portion of the lead assembly was not returned and therefore a complete evaluation could not be performed.The lead coils were broken in several locations.Scanning electron microscopy was performed, and the area had evidence of a fatigue stress fracture, fine pitting and rotational stress fracture which likely completed the fracture.Pitting and residual material were observed on the coil surface.The abraded opening found on the outer silicone tubing and the cut ends that were made during the explanted process, likely provided the path for the dried remnants of what appeared to have once been body fluids inside the outer silicone tubing.For the inner tubing fluid, there was no obvious path for fluid ingress other than the cut ends that were made during the explant process.No other anomalies were noted except the set of set screw marks observed near the end of the connector pin indicating the lead had not been fully inserted into the cavity of the generator, at one point in time.An additional pair of set screw marks provide evidence that, at one point in time, a good mechanical and electrical connection was present.
|