A siemens customer service engineer (cse) was dispatched to the customer site.After analyzing the instrument, the cse replaced reagent probe 3.The cse also replaced horizontal and vertical belts and vertical arm motors for the sample probe 2 arm.The horizontal and vertical motor for the reagent probe 3 arm was also replaced.The cse detailed the cuvette washer, cleaned the sample probe 2 and reagent probe 3 drains, verified proper air knife and vacuum at the drains, cleaned reagent openings around reagent probes 3 and 4, and cleaned the cuvette ring.The photometer, reagent probes 3 and 4, sample probes 1 and 2, and the integrated multi-sensor technology (imt) module were aligned.The barcode reader was adjusted and the sample rack vision system was aligned.Service methods testing and a quick check were run and everything passed.The cause of the discordant, falsely elevated total bilirubin results is unknown.The instrument is performing according to specifications.No further evaluation of this device is required.
|
Discrepant, falsely elevated total bilirubin results were obtained on four samples on the dimension vista 1500 instrument.The initial results were reported to the physician who questioned the results.The tests were repeated on a different dimension vista 1500 and the results were lower.The repeated results were reported to the physician.There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated total bilirubin results.
|