| Model Number |
980 |
| Medical Device Problem Code |
Device Displays Incorrect Message (2591)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
10/20/2014
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that, while in use on a patient, the 980 ventilator generated error messages.The patient was removed from the ventilator and placed on an alternate ventilator.The patient was not harmed or injured as a result of the reported event.The service engineer (se) inspected the device.The se replaced the line interface printed circuit board (pcb) and performed calibrations and extended self-testing on the device and all tests passed.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|