| Model Number |
DCJLR-A |
| Medical Device Problem Codes |
Human Factors Issue (2948); Missing Value Reason (3192)
|
| Health Effect - Clinical Code |
Laceration(s) (1946)
|
| Date of Event |
04/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This mdr submitted electronically on 04/12/2017 references accriva diagnostics' complaint number (b)(4).Method code: actual device not evaluated.Dhr review was not performed because the complaint is unrelated to product performance or packaging.Results code: no results available since no evaluation performed.Conclusion codes: human factors issue.Training deficiency.Device not returned.Accriva diagnostics has requested all data required for form 3500a.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported that an end user sustained an injury while dispensing a direct check quality control.This control is packaged in a glass ampule inside a crushable plastic dropper vial containing diluent.The end-user wore gloves and used the protective sleeve provided with the product when reconstituting the control.The purpose of the sleeve is to safeguard the end user against potential injury while handling the control.The end user had difficulty dispensing the control with the dropper vial and removed the protective sleeve.When squeezing vial even harder, the end user sustained a small cut to her right index finger which was caused by a glass shard protruding through the dropper vial.The end user washed the affected area with soap and water and applied a band.No other medical care was required and no complications were documented.
|
| |
|