• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOMET ORTHOPEDICS M2A MAGNUM SYSTEM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOMET ORTHOPEDICS M2A MAGNUM SYSTEM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Corroded (1131); Difficult to Remove (1528)
Patient Problem Reaction (2414)
Event Date 04/06/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).It is unknown if product is being returned to zimmer biomet for investigation, as its location is unknown; however, an investigation has been initiated.Once the investigation has been completed, a follow- up mdr will be submitted.
 
Event Description
Medical records received indicate difficulty removing the femoral head, which resulted in a delay of one hour.Operative report indicates the head was removed with a burr and black corrosion on the taper adapter and trunionosis were present within the joint.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of operative notes.Device history record was reviewed and no discrepancies were found.Review of the complaint history determined that no further action is required as no trends were identified.A definitive root cause could not be determined with information available.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.
 
Manufacturer Narrative
(b)(4).This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medical records and evaluation of complaint sample.Device history record (dhr) was reviewed and no discrepancies were found.Root cause was unable to be determined.M2a magnum tpr adpr ti dia42-5 0/0mmt1 item# 139256 lot# 747300 was returned and evaluated.Upon visual inspection the device had damage to the taper location and the outer shell.There was some visible discoloration to the inside taper.Sem analysis determined damage over the majority of the surface with severe corrosion attack and abundant corrosion debris.The additional information does not change the outcome of the previous investigation.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M2A MAGNUM SYSTEM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer (Section G)
BIOMET ORTHOPEDICS
56 e. bell drive
warsaw IN 46582
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6487477
MDR Text Key72653749
Report Number0001825034-2017-02493
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
PK042037
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Attorney
Type of Report Initial,Followup,Followup,Followup
Report Date 03/01/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/12/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date11/30/2016
Device Model NumberN/A
Device Catalogue Number139256
Device Lot Number747300
Other Device ID NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/17/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received02/28/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/09/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age51 YR
Patient SexMale
-
-