PENUMBRA, INC. RUBY COIL; HCG, KRD Back to Search Results
Catalog Number RBY2C0825
Medical Device Problem Codes Failure to Advance (2524); Mechanical Jam (2983)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/14/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.This report is associated with mfr report number: 3005168196-2017-00579.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure to treat a pulmonary arteriovenous malformation (pavm) using ruby coils.During the procedure, while attempting to advance a ruby coil through a non-penumbra microcatheter, the physician did not mention feeling any resistance; however, the coil became stuck half way through the microcatheter and was unable to advance any further.The ruby coil was then removed and a new ruby coil was opened.However, the same issue occurred and the ruby coil was unable to advance through the microcatheter.Therefore, the physician removed the ruby coil and completed the procedure using additional coils and the same microcatheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Results: the pet lock was intact on the proximal end of the ruby coil¿s pusher assembly.The pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.The embolization coil was intact with the pusher assembly.Conclusions: evaluation revealed that during functional analysis, both devices could be advanced through their introducer sheaths and a demonstration microcatheter without issue.The root cause of the ruby coils not being able to advance during the procedure could not be determined.Evaluation of the first ruby coil revealed the pusher assembly mid-joint was retracted proximal to the introducer sheath friction lock.This was likely incidental, and likely occurred after successful withdrawal of the ruby coil from the system.The non-penumbra microcatheter mentioned in the complaint was not returned for evaluation.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RUBY COIL
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key6487990
Report Number3005168196-2017-00578
Device Sequence Number16466059
Product Code HCG
UDI-Device Identifier00814548012971
UDI-Public00814548012971
Combination Product (Y/N)N
Initial Reporter CountryAS
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date10/20/2023
Device Catalogue NumberRBY2C0825
Device Lot NumberF66122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/14/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/14/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/12/2017
Supplement Report FDA Received Date05/26/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight90
-
-