| Model Number |
METRIX AIR |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/21/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(manufacturer narrative = t, corrected data = f) (b)(4).Returned meter evaluated with no defect found.Test strips not returned for evaluation.Most likely underlying root cause of malfunction: user's test strip had poor storage (kitchen).Test strip (b)(4).
|
| |
|
Event or Problem Description
|
|
Consumer reported complaint for high blood glucose test results.The customer is concerned with tests results from results obtained of 169 mg/dl.The customer's expected fasting blood glucose test result range is 80 to 100 mg/dl.The customer feels well and did not report any symptoms.Medical attention is not reported as a result of the actual blood glucose results.The product is not stored according to specification in the kitchen.The test strip lot manufacturer's expiration date is 12/27/2017 and open vial date is (b)(6) 2017.The meter memory was reviewed for previous test result history: (b)(6).Customer called and stated that the meter is giving his high results.Customer states he is a new diabetic and has his a1c done and obtained 6.6.
|
| |
|