| Catalog Number |
60SP050 |
| Medical Device Problem Code |
Torn Material (3024)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Customer has not returned the device to the manufacturer for device evaluation.If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
|
| |
|
Event or Problem Description
|
|
It was reported that a portex® bivona® uncuffed neonatal and pediatric tracheostomy tube had its "'ears' where you put the tape for fixation torn".The eyelets were torn after 5 days of device use.The device was replaced with a new one.No patient injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One used portex® bivona® uncuffed pediatric tracheostomy tube was returned for investigation.A visual inspection was performed and a split was observed on the flange eyelets.The most probable root cause was determined to be that the eyelet was damaged during use by coming into contact with a sharp edge, or the neck flange had a shot or a cut and it wasn't detected during the assembly process.The complaint was confirmed.
|
| |
|