• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. PORTEX® BIVONA® UNCUFFED NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. PORTEX® BIVONA® UNCUFFED NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Catalog Number 60SP050
Medical Device Problem Code Torn Material (3024)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Customer has not returned the device to the manufacturer for device evaluation.If the device becomes available and is returned and evaluated, the manufacturer will file a follow-up report detailing the results of the evaluation.
 
Event or Problem Description
It was reported that a portex® bivona® uncuffed neonatal and pediatric tracheostomy tube had its "'ears' where you put the tape for fixation torn".The eyelets were torn after 5 days of device use.The device was replaced with a new one.No patient injury was reported.
 
Additional Manufacturer Narrative
One used portex® bivona® uncuffed pediatric tracheostomy tube was returned for investigation.A visual inspection was performed and a split was observed on the flange eyelets.The most probable root cause was determined to be that the eyelet was damaged during use by coming into contact with a sharp edge, or the neck flange had a shot or a cut and it wasn't detected during the assembly process.The complaint was confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PORTEX® BIVONA® UNCUFFED NEONATAL AND PEDIATRIC TRACHEOSTOMY TUBES
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
6000 nathan lane north
minneapolis MN 55442
MDR Report Key6491305
Report Number3012307300-2017-00856
Device Sequence Number19071077
Product Code BTO
UDI-Device Identifier15021312000405
UDI-Public15021312000405
Combination Product (Y/N)N
Initial Reporter CountrySI
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor,foreign,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number60SP050
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/30/2017
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/24/2017
Supplement Date Received by Manufacturer08/08/2017
Initial Report FDA Received Date04/13/2017
Supplement Report FDA Received Date08/28/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-