Following a laparoscopic anti-reflux procedure, a patient experienced ongoing gerd leading to linx device explant.The linx device was used as part of the anti-reflux procedure.-uneventful anti-reflux procedure and device implant on (b)(6) 2015.-uneventful device explant on (b)(6) 2017 due to ongoing gerd symptoms.Device was noted as tilted downwards anteriorly and 1-2 cm below the gastroesophageal junction; investigation found that the surgeon believed he misplaced the linx device during implant.Removal surgery included a hernia repair.-after device removal, a hiatal hernia repair was performed and a second linx device (model: lxmc16, lot # 8178, serial # (b)(4)) was implanted without issue during the same surgery.
|