Model Number X SERIES
Medical Device Problem Code Loss of Power (1475)
Health Effect - Clinical Code No Information (3190)
Date of Event 03/22/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a follow-up report when our investigation is completed.
 
Event or Problem Description
Complainant alleged that while attempting to treat a patient (age & gender unknown), the device rebooted without powering off.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Additional Manufacturer Narrative
The device was returned to zoll medical corporation; the reported malfunction was observed during review of the device's activity log.The device was put through extensive testing without duplicating the reported malfunction.The device's monitor board was replaced as a precaution.The device successfully passed the final test procedure, was recertified, and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
X SERIES
Common Device Name
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
MDR Report Key6494774
Report Number1220908-2017-00792
Device Sequence Number1202341
Product Code MKJ
UDI-Device Identifier00847946017903
UDI-Public00847946017903
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K112432
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Other
Device Model NumberX SERIES
Device Catalogue NumberX SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2017
Initial Date Received by Manufacturer 03/28/2017
Supplement Date Received by Manufacturer03/28/2017
Initial Report FDA Received Date04/14/2017
Supplement Report FDA Received Date07/17/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-