• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NAVIGATION, INC. (LOUISVILLE) FUSION NAVIGATION SYSTEM; NEUROLOGICAL STEREOTAXIC INSTRUMENT Back to Search Results
Model Number FUSION
Device Problem Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Date 03/21/2017
Event Type  malfunction  
Manufacturer Narrative
No patient information provided as no patient was involved in this concern.Onsite investigation confirmed the reported event.Replacement of the system computer resolved the issue.No further issues were reported.Replaced computer was returned to manufacturer and is under analysis, no findings are available at this time.A full system check-out was completed and all tests passed.Full system functionality was confirmed and the system was returned to service.
 
Event Description
A site representative reported that there was a burning smell coming from the navigation system when they turned it on.There was no patient present when this issue was identified.
 
Manufacturer Narrative
Correction: device manufacture date updated to proper value.
 
Manufacturer Narrative
The suspect computer was returned to the manufacturer for evaluation.Hardware investigation was completed.This issue was found related to a hardware issue and was documented in a medtronic hardware anomaly tracking database.The cause of the reported issue was due to an issue with the motherboard.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FUSION NAVIGATION SYSTEM
Type of Device
NEUROLOGICAL STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer (Section G)
MEDTRONIC NAVIGATION, INC. (LOUISVILLE)
826 coal creek circle
louisville CO 80027
Manufacturer Contact
stephanie riley
attn:product quality experienc
826 coal creek circle
louisville, CO 80027-9710
7635267745
MDR Report Key6494875
MDR Text Key72878686
Report Number1723170-2017-01640
Device Sequence Number1
Product Code HAW
UDI-Device Identifier00613994380005
UDI-Public00613994380005
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K001284
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date 01/24/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFUSION
Device Catalogue Number9733560XOM
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/27/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-