| Model Number |
9240 |
| Medical Device Problem Code |
Charred (1086)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
03/13/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Returned unit was examined upon receipt.Examination showed circuit obtained damage due to an electrical fire.The fire was located in the mid-section of the circuit.Upon removing the corrugated tubing that houses the internal components, a small blackened area was noted on the wick.Further examination revealed that the wires showed signs of stress and were broken randomly.
|
| |
|
Event or Problem Description
|
|
Nurse preparing the setup of a heated anesthesia breathing circuit (part number 9240) reported that the controller display showed 70c temporarily.The circuit melted and emitted smoke.
|
| |
|