| Model Number |
48231755 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/20/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; the rod was cut.While bending the vitallium rod with the french bender the product broke in half on the back table about 2 inches in.During surgery while putting the rod in the screw the head fell off of the screw.The screw was already in the patient.All broken fragments retrieved.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual inspection, device history review, complaint history review, risk assessment; the customer reported event was confirmed via visual inspection.The screw was returned in 2 pieces.The tulip head was disengaged from the main thread body.No relevant manufacturing issues were identified as all units met specifications.The plausible root cause of the event was determined to be excessive insertion force applied during insertion leading to disengagement.
|
| |
|
Event or Problem Description
|
|
It was reported that; the rod was cut.While bending the vitallium rod with the french bender the product broke in half on the back table about 2 inches in.During surgery while putting the rod in the screw the head fell off of the screw.The screw was already in the patient.All broken fragments retrieved.
|
| |
|