| Model Number |
ETUF3214C102E |
| Medical Device Problem Code |
Leak/Splash (1354)
|
| Health Effect - Clinical Code |
Injury (2348)
|
| Date of Event |
03/28/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii aui stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm.It was reported that ct imaging discovered a type ia endoleak.The physician elected to implant an endurant aortic cuff proximal to the aui and intentionally covered the renal arteries in order to gain more seal.There was a slight proximal type i endoleak after the aortic cuff placement.The physician elected to implant eight heli-fx endoanchors and the endoleak was resolved.The physician stated that the cause of the initial type i endoleak could not be determined.The physician determined that the cause of the persistent type i endoleak, after the cuff was implanted, was due to aortic angulation.No additional clinical sequelae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Corrected information: sex, date of birth.If information is provided in the future, a supplemental report will be issued.
|
| |
|