• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ENDURANT II AUI STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ENDURANT II AUI STENT GRAFT; SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT Back to Search Results
Model Number ETUF3214C102E
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code Injury (2348)
Date of Event 03/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
An endurant ii aui stent graft system was implanted in a patient for the endovascular treatment of an abdominal aortic aneurysm.It was reported that ct imaging discovered a type ia endoleak.The physician elected to implant an endurant aortic cuff proximal to the aui and intentionally covered the renal arteries in order to gain more seal.There was a slight proximal type i endoleak after the aortic cuff placement.The physician elected to implant eight heli-fx endoanchors and the endoleak was resolved.The physician stated that the cause of the initial type i endoleak could not be determined.The physician determined that the cause of the persistent type i endoleak, after the cuff was implanted, was due to aortic angulation.No additional clinical sequelae were reported and the patient is fine.
 
Additional Manufacturer Narrative
Corrected information: sex, date of birth.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDURANT II AUI STENT GRAFT
Common Device Name
SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
MDR Report Key6498803
Report Number2953200-2017-00615
Device Sequence Number533086
Product Code MIH
UDI-Device Identifier00643169232617
UDI-Public00643169232617
Combination Product (Y/N)N
Initial Reporter StateMS
Initial Reporter CountryUS
PMA/510(K) Number
P100021
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/29/2016
Device Model NumberETUF3214C102E
Device Catalogue NumberETUF3214C102E
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 03/28/2017
Supplement Date Received by Manufacturer03/28/2017
Initial Report FDA Received Date04/17/2017
Supplement Report FDA Received Date10/02/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age69 YR
-
-