| Model Number |
ZM-541PA |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/02/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The customer reported that the transmitter is displaying different parameters than the central nurses station (cns).The heart rate on the transmitter displayed 140 beats per minute where as the cns displayed a reading of 90 beats per minute.Customer switched out the transmitter with a known working one to resolve the issue.The transmitter was sent in for evaluation.The unit was cleaned and evaluated.The reported problem of displaying other parameter was not duplicated, however our evaluation found cracking cases.Additionally, the nibp module was found to be defective.The unit has problems with the main pcb.All malfunctioning parts were replaced.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 24 hours of extended testing.Custom was sent an exchange transmitter.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the transmitter is displaying different parameters than the central nurses station (cns).
|
| |
|