• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NIHON KOHDEN CORPORATION ZM-541PA; TRANSMITTER Back to Search Results
Model Number ZM-541PA
Medical Device Problem Code Device Operates Differently Than Expected (2913)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 03/02/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The customer reported that the transmitter is displaying different parameters than the central nurses station (cns).The heart rate on the transmitter displayed 140 beats per minute where as the cns displayed a reading of 90 beats per minute.Customer switched out the transmitter with a known working one to resolve the issue.The transmitter was sent in for evaluation.The unit was cleaned and evaluated.The reported problem of displaying other parameter was not duplicated, however our evaluation found cracking cases.Additionally, the nibp module was found to be defective.The unit has problems with the main pcb.All malfunctioning parts were replaced.The unit was tested per the operator's/service manual and the results were recorded on the maintenance check sheet.The unit completed 24 hours of extended testing.Custom was sent an exchange transmitter.Nihon kohden continues to investigate the reported event.Nihon kohden will submit a supplemental report in accordance with 21 cfr section 803.56 when additional information becomes available.
 
Event or Problem Description
The customer reported that the transmitter is displaying different parameters than the central nurses station (cns).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZM-541PA
Common Device Name
TRANSMITTER
Manufacturer (Section D)
NIHON KOHDEN CORPORATION
1-31-4 nishiochia, shinjuku-ku
attn: shama mooman
tokyo, japan 161-8 560
JA  161-8560
MDR Report Key6500103
Report Number8030229-2017-00107
Device Sequence Number781977
Product Code DRT
UDI-Device Identifier04931921106617
UDI-Public04931921106617
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K043517
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 03/02/2017
Report Date (Section F) 04/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberZM-541PA
Device Catalogue NumberZM-541PA
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/13/2017
Was the Report Sent to FDA? Yes
Distributor Facility Aware Date03/02/2017
Device Age15 MO
Event Location Hospital
Date Report to Manufacturer04/18/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/18/2017
Initial Report FDA Received Date04/18/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Date Report Sent to FDA04/18/2017
-
-