• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 12F; REPLACEMENT G-TUBES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 12F; REPLACEMENT G-TUBES Back to Search Results
Model Number N/A
Health Effect - Clinical Code No Information (3190)
Date of Event 03/23/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The manufacturer has received the sample and will evaluate.Results are expected soon.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Event or Problem Description
It was reported that the balloon ruptured.
 
Additional Manufacturer Narrative
The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.The following were reviewed as part of this investigation: patient severity, frequency analysis, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the complaint of a ruptured balloon was confirmed and the cause appears to be use related.The product returned for evaluation was an 12fr tri-funnel feeding tube.The damage observed in the returned sample was characteristic of over-pressurization (burst) damage.This can occur through over-filling the feeding tube balloon.The appropriate inflation volume is listed on the inflation valve and over-inflation can be avoided by fully draining the balloon prior to inflation or re-inflation.The returned product sample was evaluated and a longitudinally-aligned split in the balloon membrane was observed.This catheter damage was determined to be typical of a burst, and the characteristics observed which supported this type of failure included: ¿ tensile weakness in the membrane at the fracture site (due to over-inflation prior to burst) ¿ granular fracture surface texture (typical of tearing failure modes) ¿ the split in the balloon was longitudinally aligned which is typical of burst-type failures, and the split traversed the entire length of the balloon.An examination of the catheter structure revealed no potential damage/defect related to manufacture of the product.These features are characteristically found due to the over-inflation and fracture of the balloon tubing.The product instruction for use (ifu) contains the following information which may be relevant to this type of event, "do not exceed the maximum recommended inflation volume as this may cause excessive pressure on the gastric mucosa and migration of the internal balloon into the peritoneal cavity".
 
Event or Problem Description
It was reported that the balloon ruptured.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARD TRI-FUNNEL REPLACEMENT GASTROSTOMY TUBE 12F
Common Device Name
REPLACEMENT G-TUBES
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key6504552
Report Number3006260740-2017-00443
Device Sequence Number19152059
Product Code KNT
UDI-Device Identifier00801741037115
UDI-Public(01)00801741037115
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K063118
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date (Section B) 05/25/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number000712
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/12/2017
Event Location Hospital
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/12/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/19/2017
Supplement Report FDA Received Date05/25/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-