The device was returned to the manufacturer for evaluation.Through visual examination of the device, a hole in the device 4 inches from the distal tip was confirmed.The nature of the hole indicated that the puncture occurred from the inside, outward.It was also noted that the working length of the device had a kink in the area where the hole was observed.The damage to the device was caused by the user and/or the use environment, and no defects in the device were noted that could have caused the observed damage.
|
It was reported that during a vascular intervention procedure to treat a lesion in the patient's superficial femoral artery (sfa), the guidewire punctured the device.Reportedly, the wire punctured the working length of the device and perforated the sfa.The injury was repaired by placing stents in the sfa.The procedure was completed and the patient was discharged as planned.
|