| Model Number |
MT22608-PNK |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
Dexcom was made aware on (b)(6) 2017 that on (b)(6) 2017, patient experienced a loss of connection between the transmitter and receiver.No additional patient or event information is available.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
The complaint device was returned for evaluation.The device was visually inspected and no defect was found.Functional testing was performed and the reported fault could not be reproduced and there were no failure detected.A pairing test was performed and the test passed.A review of the data log did not find any errors related to the customer complaint.The reported event of a loss of connection was not confirmed.A root cause could not be determined.
|
| |
|