|
It was reported that during a fenestrated endovascular aneurysm(fevar) repair of an abdominal aortic aneurysm (aaa) using a check-flo extra large introducer, the physician achieved access in the left groin.He replaced the glide wire with the lunderquist wire and introduced the 20fr sheath over the lunderquist wire.After the 20fr sheath was inserted to the required level, he attempted to remove the dilator of the 20fr sheath and the most distal portion/hub broke off.He managed to grip the tapered portion of the dilator to remove it.A section of the device did not remain inside the patient's body.It was reported that the patient did not experience any adverse effects due to this occurrence.
|
|
Investigation - evaluation: a review of documentation, instructions for use, drawings, dimension verification, complaint history, device history record, trends, manufacturing instructions, specification, quality control data, and inspection of the actual device was conducted during the investigation.During removal of the dilator from the sheath, the hub separated.The physician was able to remove the dilator by gripping on the proximal end of the dilator shaft and pull it out of the sheath.The device was returned and evaluated.The dilator and hub were separated.The dilator was able to be placed and removed from the sheath check-flo.The flare on the dilator was small.Based on the cap drawing and illustration identifying the ideal and minimum flare diameters, the flare was too small.The flare was concentric and undamaged, indicating that the flare was not deformed during both manufacturing, due to compression and torque from flla inside of cap, or during the hub separation in the procedure.This indicates that the flare was not large enough during manufacturing.Therefore, a root cause of manufacturing related / operator related was chosen.We will notify the appropriate personnel and continue to monitor for similar complaints.
|