• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIGHT SCIENCES, INC. VISCO360 VISCOSURGICAL SYSTEM; VISCOELASTIC INJECTOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIGHT SCIENCES, INC. VISCO360 VISCOSURGICAL SYSTEM; VISCOELASTIC INJECTOR Back to Search Results
Model Number 05172
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/23/2017
Event Type  malfunction  
Manufacturer Narrative
All pertinent information available to sight sciences, inc.Has been submitted.The company is submitting this mdr to ensure full compliance with 21 cfr 803.(b)(4).Submitted to fda on: april 20, 2017.
 
Event Description
The device was being used to perform viscodilation of the left eye.The microcatheter was advanced 180 degrees.Upon retraction, the surgeon felt resistance.He observed the microcatheter was severed and remained in schlemm's canal.Using retinal forceps, the surgeon successfully removed the catheter from the eye by performing 90 degree goniotomy with the microcatheter and then removing the rest of the microcatheter from the canal.The surgeon used a new device to compete the case.There were no intraoperative or post-operative complications.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISCO360 VISCOSURGICAL SYSTEM
Type of Device
VISCOELASTIC INJECTOR
Manufacturer (Section D)
SIGHT SCIENCES, INC.
3000 sand hill rd., 3-105
menlo park CA 94025
Manufacturer (Section G)
SIGHT SCIENCES, INC.
3000 sand hill rd., 3-105
menlo park CA 94025
Manufacturer Contact
anne-marie ripley
3000 sand hill rd., 3-105
menlo park, CA 94025
6616458546
MDR Report Key6508202
MDR Text Key73460066
Report Number3010363671-2017-00017
Device Sequence Number1
Product Code MRH
UDI-Device Identifier00858027006020
UDI-Public(01)00858027006020(17)171231(10)1005123
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K143205
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Physician
Type of Report Initial
Report Date 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date12/31/2017
Device Model Number05172
Device Catalogue Number05172
Device Lot Number1005123
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/28/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/12/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
-
-