| Model Number |
33422 |
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that the rubber was cracking around the male end that connects to the urinary bag.
|
| |
|
Additional Manufacturer Narrative
|
|
Received 1 unopened silastic foley catheter.The reported event was confirmed as use related.The sample was examined and found what may be evidence of ozone damage.Factors which may contribute to this problem are storage in excessive heat or cold, or storage near fluorescent lighting, electric motors or ozone generating equipment (i.E., x-ray equipment).For this reason, latex products should not be stored in close proximity to any of these electrical devices or equipment.Such products should be stored in their original cartons with the end flaps closed, away from direct lighting and extreme temperatures and should be rotated to ensure that the oldest products are used first.The reported condition could not be attributed to a manufacturing-related process.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "visually inspect the product for any imperfections or surface deterioration prior to use." (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the rubber was cracking around the male end that connects to the urinary bag.
|
| |
|
Search Alerts/Recalls
|