| Model Number |
13 |
| Medical Device Problem Code |
Material Deformation (2976)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/22/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Ra has received the incident device and the product has been assigned to engineering for evaluation.A follow-up report will be sent upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Event description: (on (b)(6) 2017: during a preparation), a medical staff tried to attach the insert g4550 to a clamp, however; it did not fit.Another package of insert was opened but the medical staff confirmed deformation at the button.Thus the medical staff canceled to use the second one.This event didn't affect the patient.Additional information received via email from distributor april 5, 2017 - the defects for the g-4550 were noticed before the procedure.Procedure performed: unknown.Type of intervention: na.Patient status: no health damage was reported for the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Investigation summary: the event unit was returned for evaluation.Upon visual inspection, engineering determined that the male attachment features (buttons) of the insert were damaged, which would affect the unit's ability to attach to the clamp.It is likely that the root cause of the event was improper technique during installation of the insert, as this can cause damage to the attachment features of the insert and make it unable to remain attached to the clamp.The instructions for use (ifu) states, "to place inserts, open the clamp to spread both clamp jaws.Place the button located on the proximal end of the insert into the mounting hole.Snap the other button into the clamp by pressing forward and down on the insert with firm hand pressure.Do not press down directly on this button."applied medical will continue to monitor its vigilance system for trends and take appropriate actions as necessary to ensure the performance and safety of its products.
|
| |
|
Search Alerts/Recalls
|