• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SYNTHES MONUMENT 10MM/130 DEG TI CANN TFNA 340MM/RIGHT ¿ STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SYNTHES MONUMENT 10MM/130 DEG TI CANN TFNA 340MM/RIGHT ¿ STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Catalog Number 04.037.054S
Device Problem Break (1069)
Patient Problems Pain (1994); Ambulation Difficulties (2544)
Event Type  Injury  
Manufacturer Narrative
(b)(4).Manufacturing location: (b)(4); manufacturing date: 23-oct-2015; expiration date: 30-sep-2025; part #: 04.037.054s, lot#: 9926305 (sterile), component parts reviewed: no ncrs were generated during production.Review of the device history record(s) showed that there were no issues during the manufacture of the product that would contribute to this complaint condition.¿sterility documentation was reviewed and determined to be conforming.¿ the returned nail was received along with the concomitant helical blade (04.038.285) which was used during the initial procedure; since the blade was concomitant no investigation will be completed for it.The returned 10mm 130° ti trochanteric fixation nail advanced (tfna) nail (04.037.054s, 9926305) is part of tfna systems (dsus/trm/0815/0682, dsem/trm/0915/0523, dsem/trm/0514/0052) and is used for intramedullary fixation of proximal femoral fractures.The returned nail was received broken in the region where the helical blade was inserted; also both the returned nail and blade were received with superficial damage indicative of extraction.Replication of the complaint condition is inapplicable since the complaint condition was confirmed due to the returned nail being received broken.As part of this investigation a visual inspection, drawing review, root cause analysis, and risk management assessment were performed.The returned nail (04.037.054s, 9926305) was manufactured on 23oct15 and drawings were reviewed and determined to be suitable for the intended design, application, and dimensional conformity when used and handled as recommended.No mrrs, ncrs, or actions related to the complaint condition were generated during production of the returned part(s).Review of device history record(s) showed that there were no issues during the manufacture of the product(s) which would contribute to the complaint condition.During the investigation no unidentified product design issues or discrepancies were observed that may have contributed to the complaint condition.Based on the available details it is not possible to determine a definitive root cause for the complaint condition; however it was noted that the patient had femur cancer and osteopenia, which could have contributed to inadequate healing and subsequently nail breakage.It was also stated that a bigger nail/helical blade was used during the revision surgery, so it is possible that the initial implants were undersized and contributed to the nail failure.During the investigation no unidentified product design issues or discrepancies were observed that may have contributed to the complaint condition.The associated dcrm adequately addresses the complaint condition and the calculated occurrence rate does not exceed established limits.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the patient underwent revision surgery on (b)(6) 2017 due to a broken nail.The patient was originally implanted with a right trochanteric femoral nail-advanced (tfna) nail and helical blade on (b)(6) 2016.Original surgery was completed successfully without complications.The patient reported that she experienced pain for nine months and felt like she couldn¿t walk.Upon follow-up with her surgeon, it was determined by x-ray taken on (b)(6) 2017 that the nail was broken.Patient was returned to surgery on (b)(6) 2017 where the nail and helical blade were removed.Patient was revised to a larger nail and helical blade.Revision surgery was completed successfully without surgical delay.It is reported that, two (2) months post-operative, patient is able to walk.Concomitant medical products: helical blade (part 04.038.285, lot number unknown, quantity 1).This report is for one (1) tfna nail.This is report 1 of 1 for (b)(4).
 
Manufacturer Narrative
Additional patient identifier reported as (b)(6).Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Patient underwent a right femoral surgery on (b)(6) 2016.Patient experienced excruciating pain in the mid right thigh and she couldn't even stand up around (b)(6) 2017.Postoperative x-ray taken on (b)(6) 2017 revealed that the nail had completely broken in the region where the helical blade was inserted.On (b)(6) 2017, the patient was revised to a 340 mm synthes rod and 12 mm thickness, 85 mm triflange nail, 36 distal cortical screw.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
10MM/130 DEG TI CANN TFNA 340MM/RIGHT ¿ STERILE
Type of Device
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer (Section G)
SYNTHES MONUMENT
1051 synthes ave
monument CO 80132
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6513689
MDR Text Key73449817
Report Number1719045-2017-10335
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K131548
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/22/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number04.037.054S
Device Lot Number9926305
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/07/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received08/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/23/2015
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
HELICAL BLADE(04.038.285, LOT UNKNOWN, QTY 1)
Patient Outcome(s) Required Intervention;
Patient Weight75
-
-