Model Number ZXT150
Medical Device Problem Code Unintended Movement (3026)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 01/10/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a toric intraocular lens (iol) was implanted without complications, in the left eye of a male patient.Six weeks post operation, it was noted that the lens was dislocated.Reportedly, the lens was repositioned in a secondary surgical procedure on (b)(6) 2017.The lens was reported being perfectly centered.The patient tolerated the procedure well, and was returned to the recovery room in good condition.Additional information was received and it was learnt that the problem first occurred on (b)(6) 2017; the day the lens was implanted.The lens kept falling off the capsular bag, the iol continued to move and had moved twice and on the third time the lens was explanted and replaced.The lens was explanted on (b)(6) 2017.There was no incision enlargement, no vitrectomy was performed and no patient injury was reported.The patient was reported doing well.The lens was replaced with a different lens model, z9002, and smaller dioptre, +16.0.No further information was provided.
 
Additional Manufacturer Narrative
Device evaluation: the intraocular lens (iol) was not returned at the manufacturing site; therefore product testing could not be performed and the customer's reported complaint could not be verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search revealed that no similar complaints for this order number have been received.Labeling review: the directions for use (dfu) were reviewed.The directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS SYMFONY TORIC
Common Device Name
MULTIFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6514045
Report Number9614546-2017-00315
Device Sequence Number16167461
Product Code POE
UDI-Device Identifier05050474605152
UDI-Public(01)05050474605152(17)211102
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/12/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date11/02/2021
Device Model NumberZXT150
Device Catalogue NumberZXT150U175
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/27/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/22/2017
Supplement Report FDA Received Date06/12/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-