| Model Number |
4085-9 |
| Medical Device Problem Code |
Incorrect Measurement (1383)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Attempts have been made to obtain the product.The product has not been returned to masimo to allow an analysis to be performed.If the product is returned for evaluation or new information is obtained, a follow up report will be submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that there were broken rd ge cables and questionable spo2 values.Returning 6 rd ge 4085 cables.Exposed wires at the sensor connection.No consequences or impact to patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned cable was evaluated.During evaluation the cable passed all visual and functional testing.The cable was determined to be functioning as designed.A service history record review reveals that this unit was in the field for over eleven (11) months with no previous reported issues related to this reported event.Corrected data: model # from "4085" to "4085-9".
|
| |
|