• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEXCOM, INC. DEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM Back to Search Results
Model Number 9500-27
Device Problems Use of Device Problem (1670); Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4). The g5 system is associated with product code pqf.
 
Event Description
Dexcom was made aware on (b)(6) 2017, that on (b)(6) 2017, there was a continuous glucose monitoring (cgm) inaccuracy compared to blood glucose (bg) meter that occurred. The sensor was inserted on the arm, on (b)(6) 2017. No additional event or patient information is available. Labeling indicates: do not insert the sensor component of the dexcom g5 mobile system in a site other than the belly/abdomen (ages 2 years and older) or the upper buttocks (ages 2 to 17 years). The placement and insertion of the sensor component of the dexcom g5 mobile system is not approved for other sites. If placed in other areas, the dexcom g5 mobile system may not function properly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameDEXCOM G5 MOBILE CONTINUOUS GLUCOSE MONITORING SYSTEM
Type of DeviceCONTINUOUS GLUCOSE MONITOR
Manufacturer (Section D)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer (Section G)
DEXCOM, INC.
6340 sequence drive
san diego CA 92121
Manufacturer Contact
kipp durbin
6340 sequence drive
san diego, CA 92121
8582000200
MDR Report Key6516811
MDR Text Key73566283
Report Number3004753838-2017-33214
Device Sequence Number1
Product Code MDS
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P120005
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial
Report Date 03/29/2017
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received04/24/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number9500-27
Device Catalogue NumberSTS-GL-011
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received03/29/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-