NEWPORT MEDICAL INSTRUMENTS, INC. E360 VENTILATOR; VENTILATOR, CONTINUOUS, FACILITY USE Back to Search Results
Model Number E360
Medical Device Problem Code Failure to Deliver (2338)
Health Effect - Clinical Code No Information (3190)
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that there were no air or oxygen readings.Patient involvement information was not provided by the customer.The ventilator was evaluated by a non-medtronic service provider and the oxygen delivery was out of specification.The oxygen sensor was replaced, which resolved the condition.Medtronic was not authorized to service the ventilator.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
E360 VENTILATOR
Common Device Name
VENTILATOR, CONTINUOUS, FACILITY USE
Manufacturer (Section D)
NEWPORT MEDICAL INSTRUMENTS, INC.
1620 sunflower ave
costa mesa CA 92626
MDR Report Key6517360
Report Number2023050-2017-05210
Device Sequence Number11095543
Product Code CBK
UDI-Device Identifier10884521540989
UDI-Public10884521540989
Combination Product (Y/N)N
Initial Reporter CountryIN
PMA/510(K) Number
K053502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 03/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberE360
Device Catalogue NumberE360E-WW-IN
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 03/28/2017
Initial Report FDA Received Date04/24/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No