| Model Number |
482802645 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; the extraction surgery was performed due to healing of the disease.During the extraction surgery, the breakage of l5 right side screw was found.The tip of the screw was remained to the patient and the surgery was finished.
|
| |
|
Additional Manufacturer Narrative
|
|
Method: visual inspection, device history review, complaint history review, risk assessment.Result: manufacturing records were reviewed for the corresponding lot and no relevant issues were identified.The returned screw was inspected and it was confirmed to be fractured.Conclusion: the plausible root cause of the event is high activity of the patient leading to stress on the implant causing fracture.
|
| |
|
Event or Problem Description
|
|
It was reported that; the extraction surgery was performed due to healing of the disease.During the extraction surgery, the breakage of l5 right side screw was found.The tip of the screw was remained to the patient and the surgery was finished.
|
| |
|