• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL, SAGITTAL SAW POWER; SAW, POWERED, AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL, SAGITTAL SAW POWER; SAW, POWERED, AND ACCESSORIES Back to Search Results
Catalog Number 7205786S
Device Problem Device Operates Differently Than Expected (2913)
Patient Problem No Information (3190)
Event Date 03/29/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during a procedure the device was not working.A backup device was not available.No additional information was provided.
 
Manufacturer Narrative
A visual inspection was performed on the product and no issue was observed.A functional evaluation revealed the head assembly was making a noise and sticking.Disassembly revealed the instrument had a cracked lohet and the threads on the body were damaged.The complaint was confirmed and the root cause has been determined to be expected wear/tear due to usage and time in the field.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL, SAGITTAL SAW POWER
Type of Device
SAW, POWERED, AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer (Section G)
SMITH & NEPHEW, INC.
150 minuteman road
andover MA 01810
Manufacturer Contact
jim gonzales
7000 w william cannon dr.
austin, TX 78735
MDR Report Key6518736
MDR Text Key73813186
Report Number3003604053-2017-00050
Device Sequence Number1
Product Code HAB
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 10/04/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/25/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number7205786S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/27/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-