• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX I 100BX

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 ROCHESTER MEC ULTRAFLEX I 100BX Back to Search Results
Model Number 33103
Health Effect - Clinical Code Injury (2348)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.
 
Event or Problem Description
It was reported that the patient developed sores on his penis.
 
Additional Manufacturer Narrative
Received 1 opened ultraflex mec.The sample was received with the package opened , therefore it could not be evaluated since the integrity of the test would be compromised and the results may not be accurate.The reported complaint was found to be inconclusive due to the way the sample was received.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use state the following: "description/indication the self-adhering male external catheter is designed for the management of male urinary incontinence.Contraindication do not use on irritated or compromised skin.Precaution do not use if allergic reaction occurs.For good hygiene, change catheter daily.Use of a single device for longer periods than 24 hours may increase the risk of complications.Directions: to apply 1) wash penis with mild soap and warm water.Dry thoroughly.2) trim pubic hair if necessary.3) unroll self-adhering catheter over penis.4) gently squeeze the catheter to properly seal adhesive to the skin.Important: wear time may be significantly reduced if adhesive is not properly sealed to the skin.5) connect to drainage device.Directions: to remove gently roll catheter off the penis.Note: if necessary, apply a warm wet compress (such as a wet washcloth) around the catheter to help loosen the adhesive." (b)(4).
 
Event or Problem Description
It was reported that the patient developed sores on his penis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROCHESTER MEC ULTRAFLEX I 100BX
Common Device Name
MEC
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
parque industrial aeropuerto
covington GA 30014
MDR Report Key6518803
Report Number1018233-2017-02032
Device Sequence Number736620
Product Code NNX
UDI-Device Identifier00801741070686
UDI-Public(01)00801741070686(17)210928(10)JUAX0330
Combination Product (Y/N)N
Initial Reporter StateMA
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer,other
Initial Reporter Occupation Patient
Type of Report Initial,Followup
Report Date (Section B) 02/19/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Lay User/Patient
Device Expiration Date09/28/2021
Device Model Number33103
Device Catalogue Number33103
Device Lot NumberJUAX0330
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/03/2017
Initial Date Received by Manufacturer 04/03/2017
Supplement Date Received by Manufacturer02/14/2018
Initial Report FDA Received Date04/25/2017
Supplement Report FDA Received Date02/19/2018
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
-
-