• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS TECNIS; TORIC IOLS Back to Search Results
Model Number ZCT450
Medical Device Problem Code Incorrect Measurement (1383)
Health Effect - Clinical Code No Code Available (3191)
Date of Event 03/28/2017
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).All pertinent information available to the manufacturer has been submitted.
 
Event or Problem Description
It was reported that a zct450 20.0 diopter intraocular lens (iol) was explanted due from a male patient's operative eye due to lens calculations over corrected.The lens was replaced with a model zct300 16.0 diopter iol.There was no incision enlargement, no vitrectomy and no capsule tear.The patient is recovering well.
 
Additional Manufacturer Narrative
Device available for evaluation ¿ yes, returned to manufacturer on 07/07/2017.Device returned to manufacturer ¿ yes.Device evaluation: the intraocular lens (iol) was returned to the manufacturing site for evaluation.Visual inspection with the unaided eye shows the product is cut in pieces, most probably to make explant possible.Considering the condition of the lens additional analysis is not possible.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.A search on complaints revealed that no similar complaints have been received for this production order number.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.As a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
TORIC IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6521307
Report Number9614546-2017-00325
Device Sequence Number6333
Product Code HQL
UDI-Device Identifier05050474602984
UDI-Public(01)05050474602984(17)201028
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Device Explanted Year2017
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/16/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Expiration Date10/28/2020
Device Model NumberZCT450
Device Catalogue NumberZCT450U200
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2017
Initial Date Received by Manufacturer 03/29/2017
Supplement Date Received by Manufacturer07/23/2017
Initial Report FDA Received Date04/26/2017
Supplement Report FDA Received Date08/16/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age88 YR
-
-