The cause for the discordant advia centaur xp ca19-9 results is possibly due to the differences in assay methods.The ifu states in the warning before the intended use section: "warning: the concentration of ca 19-9 in a given specimen, as determined by assays from different manufacturers, can vary due to differences in assay methods and reagent specificity.The results reported by the laboratory to the physician must include the identity of the assay for ca 19-9 used.Values obtained with different assay methods cannot be used interchangeably.If, in the course of monitoring a patient, the assay method used for determining serial levels of ca 19-9 is changed, the laboratory must perform additional serial testing to confirm baseline values.The advia centaur ca 19-9 assay is based on the 1116-ns-19-9 antibody available through agreement with (b)(4) assays using antibodies other than 1116-ns-19-9 may give different results.Both the advia centaur xp ca 19-9 results and the alternate laboratory results are clearly elevated indicating the presence of malignant disease.Based on the available information, the advia centaur xp ca 19-9 assay is performing as intended.The instrument is performing within specifications.No further evaluation of the device is required.The ifu states in the interpretation of results section: "results of this assay should always be interpreted in conjunction with patient's medical history, clinical presentation and other findings.".
|