| Catalog Number |
FEL06080 |
| Medical Device Problem Codes |
Fracture (1260); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/23/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device history records are being reviewed.The event is currently under investigation.The information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant was unable to provide any further patient or procedural details to bard.
|
| |
|
Event or Problem Description
|
|
It was reported that during a fistulagram procedure, the sheath of the delivery system broke at the hub while being deployed through a 9 f introducer sheath.Reportedly, the stent graft could be implanted successfully.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The lot history records have been reviewed with special attention to the manufacturing and inspection of this product and the product was found to have met all specifications prior to shipment.As part of all complaint investigations, an attempt was made to evaluate the subject device and the device usage.In this case, no sample or image was provided for evaluation.Therefore, the reported event could not be reproduced.Potential contributing factors to the reported event have been considered.Previous investigations of similar complaints have been reviewed.This type of event may be associated with a difficult vessel anatomy leading to increased friction and subsequent sheath fracture.Insufficient flushing of the device may be another potential contributing factor.In this case, no additional information regarding the vessel anatomy, the stent placement site or the procedure performed was provided.On the basis of the information available and as no sample was returned for evaluation, a definite root cause for the reported failure could not be determined.The ifu states that the device must be flushed with sterile saline.Also the ifu states: "the safety and effectiveness of the device when placed across a tight bend (.) has not been evaluated.Prior to stent graft deployment, ensure that the proximal (inflow) stent graft end is positioned in a straight section of the lumen to reduce the risk of higher deployment forces and possible endovascular system failure." and "do not kink the delivery catheter or use excessive force during delivery to the target lesion." also the ifu states: "if unusual resistance or high deployment force is encountered during stent graft deployment, abort the procedure, remove the delivery system and use an alternative device.".
|
| |
|
Event or Problem Description
|
|
It was reported that during a fistulagram procedure, the sheath of the delivery system broke at the hub while being deployed through a 9 f introducer sheath.Reportedly, the stent graft could be implanted successfully.There was no reported patient injury.
|
| |
|