• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. TURON SHOULDER; HEAD, HUMERAL, NEUTRAL, 46-16

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. TURON SHOULDER; HEAD, HUMERAL, NEUTRAL, 46-16 Back to Search Results
Catalog Number 520-46-016
Health Effect - Clinical Codes Pain (1994); Discomfort (2330)
Date of Event 03/27/2017
Type of Reportable Event Serious Injury
Event or Problem Description
Revision surgery - overstuffing of the joint due to too large of a humeral head.The patient was suffering from pain and discomfort.
 
Additional Manufacturer Narrative
The reason for this revision surgery was pain and discomfort.The in-vivo length of patient service for the implant was 2.5 years.There is no information in this complaint about any patient injuries, activities, or accidents that may have contributed to the need for this revision surgery.There are no reported pre-existing patient health conditions.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history records (dhrs) revealed no discrepancies or issues with the manufacturing of this part.All critical dimensions, design criteria and specifications in effect at the time the part was manufactured were met.The product complaint report history was reviewed and no trends or on-going issues were deemed as present or in need of review.This event is deemed as non-product related.The root cause for this event was an overstuffed patient joint, due to a too large humeral head, resulting in patient pain and discomfort.Inventory containment is not required since there are no indications of a product or process issue affecting implant safety or effectiveness.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TURON SHOULDER
Common Device Name
HEAD, HUMERAL, NEUTRAL, 46-16
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd.
austin TX 78758 5445
MDR Report Key6522697
Report Number1644408-2017-00281
Device Sequence Number10946318
Product Code KWS
UDI-Device Identifier00888912145145
UDI-Public(01)00888912145145
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K080402
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2014
Device Explanted Year2017
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date06/30/2019
Device Catalogue Number520-46-016
Device Lot Number917C1038
Was Device Available for Evaluation? No
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 03/27/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date04/26/2017
Supplement Report FDA Received Date05/31/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
520-00-001,LOT 195F1038; 521-01-246,LOT 192G1075
Outcome Attributed to Adverse Event Required Intervention;
Patient Age54 YR
Patient Weight70
-
-