| Model Number |
25-878-07-91 |
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
03/27/2017
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
When the doctor was putting the plate on to the segmentalized fibula bone after pre-drilling the holes, the 7 mm screws partially threaded before becoming stuck in the bone.Increased force on the screwdriver to mobilize screws caused the screw heads to strip and they became unusable.Doctor was forced to remove the screws with a wire-driver, and one of the screw heads broke off from the threads.The doctor had to move the plate to a non-ideal location slightly below the previous screw holes in order to fixate the plate while avoiding the stripped screw body.Patient is doing well and no other adverse effects from this incident have been reported at this point.
|
| |
|
Additional Manufacturer Narrative
|
|
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage was calculated and it is determined that the complaint percentage falls within the design risk limits adhered to at klm.The review of the device history records was not possible due to no lot number being provided.The failure root cause cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
|
| |
|