| Catalog Number |
65P050 |
| Medical Device Problem Code |
Air Leak (1008)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
03/31/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Smiths medical has received the sample device.A full evaluation is anticipated, but not yet begun as the device is currently in transit to the investigation site.Smiths medical will file a follow-up report detailing the results of the evaluation once it is completed.Additional 510k number - k083641.
|
| |
|
Event or Problem Description
|
|
It was reported that during patient use of the bivona® aire-cuf® neonatal and pediatric tracheostomy tube, the device cuff had an air leak and was unable to remain inflated.The tracheostomy tube was replaced after the reported issue was observed.It is unknown how long the device was in patient use before the incident was observed.The device cuff patency was tested prior to patient use.The event was considered resolved.No permanent injury was reported.
|
| |
|
Additional Manufacturer Narrative
|
|
One bivona® 5.0 mm aire-cuf® neonatal and pediatric tracheostomy tube was returned for investigation.Visual inspection revealed no defects or abnormalities with the returned device.During functional testing, the device cuff was inflated with air and the device was submerged under water.No bubbles (indicating a leak) were observed escaping from the device.Investigation was unable to confirm the reported event and found that the device functioned as intended.
|
| |
|