| Catalog Number |
482319550 |
| Medical Device Problem Codes |
Detachment Of Device Component (1104); Mechanical Problem (1384)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); Device Embedded In Tissue or Plaque (3165)
|
| Date of Event |
03/29/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that; dr.Placed 9.5x50 screw in left s1 pedicle during revision surgery.Dr was replacing current 8.5 diameter screw with larger due to having a 'loose feel'.Upon placement, dr wanted to removed and redirect due to 'poor' trajectory, but could not.Dr.Broke handle/driver/nut trying to remove with primary driver, then switch to lumbar removal driver over a locked down piece of rod with blocker to top of tulip to try to 'helicopter' screw out since it was so stuck.During process of trying to remove, he broke tulip/ head off of screw off of main screw shaft.Dr could not remove remainder of broken screw from patient with lumbar removal set - so he left it in place due to not having effect pm patient's outcome and moved on with case.
|
| |
|
Additional Manufacturer Narrative
|
|
Risk assessment; visual, dimensional and functional analysis could not be performed as the device was not returned and no lot# was provided.The most likely root cause is excessive force was applied to the screw causing it to disengage, with incorrect trajectory of the screw also contributing to the event.
|
| |
|
Event or Problem Description
|
|
It was reported that; dr.Placed 9.5x50 screw in left s1 pedicle during revision surgery.Dr was replacing current 8.5 diameter screw with larger due to having a 'loose feel'.Upon placement, dr wanted to removed and redirect due to 'poor' trajectory, but could not.Dr.Broke handle/driver/nut trying to remove with primary driver, then switch to lumbar removal driver over a locked down piece of rod with blocker to top of tulip to try to 'helicopter' screw out since it was so stuck.During process of trying to remove, he broketulip/ head off of screw off of main screw shaft.Dr could not remove remainder of broken screw from patient with lumbar removal set - so he left it in place due to not having effect pm patient's outcome and moved on with case.
|
| |
|