| Model Number |
N/A |
| Medical Device Problem Code |
Malposition of Device (2616)
|
| Health Effect - Clinical Code |
No Information (3190)
|
| Date of Event |
03/24/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Exemption number e2016018 b.Braun medical, inc.(importer) is submitting this report on behalf of b.Braun melsungen ag (manufacturer).This report has been identified as b.Braun melsungen ag internal report # (b)(4).We received one used capillary housing of an introcan safety 3-w pur 20g 1.1x32mm-eu in open packaging.The needle inclusive housing was not handed over by the customer.The returned sample was taken to a visual inspection.The capillary is cut off approximately 10 mm away from the capillary housing.The cut off portion is missing and was not delivered by the customer.The cut-off point of the capillary shows the typical v-shaped cut (e.G.If damaged by the tip of the needle).We assume that this puncture / cut off was caused by withdrawing the introcan cannula and pushing forward again.According to the damage at the used sample we assume of an application error (application error; see instruction for use: never reinsert the needle inside the catheter once the needle has been partially or completely withdrawn as it may pierce and/or sever the catheter).Consequently, we consider the complaint not confirmed.Device history record review (dhr) :- reviewed the device history record and no abnormalities found during in process and final control inspection.
|
| |
|
Event or Problem Description
|
|
As reported by the user facility (translation of user facility information by bbm sales organization in united kingdom): plastic tube which sits in vein snapped.
|
| |
|