Model Number 25-092-08-91
Health Effect - Clinical Code Device Embedded In Tissue or Plaque (3165)
Date of Event 03/29/2017
Type of Reportable Event Serious Injury
Event or Problem Description
As they were removing the screws the tip of the screw broke off in the patient.The broken piece of screw was left in the patient's maxilla.It was discovered in the post op x-ray after the case was completed.
 
Additional Manufacturer Narrative
An investigation was performed on the basis of complaint statistics as no device was returned for evaluation.The complaint percentage falls within the design risk limits adhered to at klm.The review of the product device history records was performed on the lot number provided.In this investigation no anomalies were discovered.The root cause for the failure cannot be determined due to no device being returned.If further information is obtained that might add value to the contents of the investigation report, an additional follow-up report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEVEL ONE CMF
Common Device Name
SCREW
Manufacturer (Section D)
KARL LEIBINGER MEDIZINTECHNIK GMBH & CO. KG
kolbinger strabe 10
muhlheim/donau, 78570
GM  78570
MDR Report Key6527507
Report Number9610905-2017-00039
Device Sequence Number13785779
Product Code DZL
UDI-Device Identifier00888118037206
UDI-Public(01)00888118037206
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K042573
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2017
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 03/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Physician
Device Model Number25-092-08-91
Device Lot Number33180604
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 03/29/2017
Supplement Date Received by Manufacturer03/29/2017
Initial Report FDA Received Date04/27/2017
Supplement Report FDA Received Date09/13/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-