(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The armada 14 referenced in describe event or problem is filed under a separate medwatch report number.Evaluation summary: visual inspections were performed on the returned device.The stent movement was confirmed.The investigation determined the reported difficulties appear to be related to operational context.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other similar incidents.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
|