AV-TEMECULA-CT OMNILINK ELITE PERIPHERAL STENT SYSTEM Back to Search Results
Catalog Number 1012625-39
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The armada 14 referenced in describe event or problem is filed under a separate medwatch report number.Evaluation summary: visual inspections were performed on the returned device.The stent movement was confirmed.The investigation determined the reported difficulties appear to be related to operational context.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history of the reported lot revealed no other similar incidents.There is no indication of a product quality issue with respect to the design, manufacture, or labeling of the device.
 
Event or Problem Description
It was reported that the procedure was to treat a lesion located in the left common iliac artery.When the over the wire omni elite 9.0 mm/39 mm/80 cm stent was opened, the stent was noted to be loose on the balloon.The device was not used and there was no patient involvement.Another omni link stent was used as replacement.When the armada 14 pta balloon catheter was being used for dilatation the balloon ruptured below rated burst pressure.The device was prepared according to the instructions for use and another armada 14 balloon was used to successfully complete the procedure.There were no reported adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OMNILINK ELITE PERIPHERAL STENT SYSTEM
Common Device Name
PERIPHERAL STENT SYSTEM
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
MDR Report Key6528597
Report Number2024168-2017-03718
Device Sequence Number324623
Product Code NIO
UDI-Device Identifier08717648179341
UDI-Public(01)08717648179341(17)180331(10)4102941
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P110043
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Health Professional
Remedial Action Other
Type of Report Initial
Report Date (Section B) 04/27/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/31/2018
Device Catalogue Number1012625-39
Device Lot Number4102941
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/11/2017
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/03/2017
Initial Report FDA Received Date04/27/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
-
-