Catalog Number 42544
Medical Device Problem Code Inaccurate Flow Rate (1249)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/28/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product testing is in progress.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported to medtronic neurosurgery that the reservoir's flexibility was found to be poor at the pre-operation test.According to the report, the reservoir could not be easily filled.
 
Additional Manufacturer Narrative
The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation and leak testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.
 
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CSF-FLOW CONTROL VALVE, BURR HOLE, MEDIUM
Common Device Name
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
MDR Report Key6529360
Report Number2021898-2017-00235
Device Sequence Number11099023
Product Code JXG
UDI-Device Identifier00643169466913
UDI-Public00643169466913
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K831678
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Health Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 03/29/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2020
Device Catalogue Number42544
Device Lot NumberD84234
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/25/2017
Initial Date Received by Manufacturer 03/29/2017
Supplement Date Received by ManufacturerNot provided
05/05/2017
Initial Report FDA Received Date04/28/2017
Supplement Report FDA Received Date06/02/2017
10/03/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-