| Catalog Number |
42544 |
| Medical Device Problem Code |
Inaccurate Flow Rate (1249)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
03/28/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The product testing is in progress.A review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported to medtronic neurosurgery that the reservoir's flexibility was found to be poor at the pre-operation test.According to the report, the reservoir could not be easily filled.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned valve was patent.It met the requirements for reflux, pressure-flow, preimplantation and leak testing.There was proteinaceous debris observed within the interior and exterior of the valve.Debris within the valve may hold pressure-flow controlling mechanisms open affecting the flow of fluid through the valve.The instructions for use (ifu) that accompany the device caution that ¿shunt obstruction may occur in any of the components of the shunt system.The system may become occluded internally due to tissue fragments, blood clots, tumor cell aggregates, bacterial colonization, or other debris.¿ a review of the manufacturing records showed no anomalies.All valves are 100% tested at the time of manufacture.
|
| |
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|