• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT LABORATORIES ARCHITECT CA 19-9XR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT LABORATORIES ARCHITECT CA 19-9XR Back to Search Results
Catalog Number 02K91-38
Device Problem Low Test Results (2458)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, a search for similar complaints, review of instrument logs, a review of labeling and accuracy testing.No adverse trend was identified for the customer's issue.Labeling was reviewed and found to be adequate.The product was not available for return.An internal panel was tested with retained kits of the likely cause reagent lot and accuracy testing met all specifications.A malfunction was identified as the device failed to meet performance specifications or otherwise perform as intended at the customer site.Based on all available information and abbott diagnostics' complaint investigation no product deficiency was identified.
 
Event Description
The customer observed falsely depressed ca19-9 results on the architect i2000sr analyzer.The following data was provided: initial 26.8, repeat 11.4.There was no impact to patient management reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT CA 19-9XR
Type of Device
CA 19-9
Manufacturer (Section D)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer (Section G)
ABBOTT LABORATORIES
100 abbott park road
abbott park IL 60064 3500
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6534852
MDR Text Key74167077
Report Number1415939-2017-00082
Device Sequence Number1
Product Code NIG
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K052000
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial
Report Date 05/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/23/2016
Device Catalogue Number02K91-38
Device Lot Number50405M500
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 04/18/2017
Initial Date FDA Received05/01/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
ARCHITECT I2000SR ANALYZER; LN 03M74-01 SN (B)(4)
-
-