Model Number ZA9003
Medical Device Problem Code Scratched Material (3020)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 03/29/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Implant and explant dates: if implanted or explanted, give date: not applicable no patient contact reported.All pertinent information available to abbott medical optics has been submitted.
 
Event or Problem Description
It was reported that a scratch was noticed on an intraocular lens (iol) upon opening the lens case.No patient contact was reported.No further information was provided.
 
Additional Manufacturer Narrative
Device available for evaluation? yes.Returned to manufacturer on: 05/15/2017.Device returned to manufacturer? yes.Device evaluation: the intraocular lens (iol) was returned at the manufacturing site for evaluation.Visual inspection at 10x microscope magnification showed viscoelastic residue on the lens, indicating that the unit was handled and prepared for use.No damage was observed on the lens.The customer's reported complaint was not verified.Manufacturing records review: the manufacturing records for the intraocular lens were reviewed.During manufacturing the lenses are 100% cleaned and inspected at 10x microscope magnification.The product was manufactured and released according to specification.A search revealed that no additional complaints for this order number have been received.Labeling review: the directions for use (dfu) adequately provides instructions and precautions along with warnings for the proper use and handling of the device.The dfu states to examine the lens thoroughly to ensure particles have not become attached to it, and examine the lens optical surfaces for other defects.As a result of the investigation there is no indication of a product quality deficiency and the reported issue was not verified.All pertinent information available to abbott medical optics has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
MDR Report Key6535825
Report Number2648035-2017-00764
Device Sequence Number943515
Product Code HQL
UDI-Device Identifier05050474529137
UDI-Public(01)05050474529137(17)180717
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
P990080
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/09/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Physician
Device Expiration Date07/17/2018
Device Model NumberZA9003
Device Catalogue NumberZA90030260
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/15/2017
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/03/2017
Supplement Date Received by ManufacturerNot provided
Initial Report FDA Received Date05/01/2017
Supplement Report FDA Received Date06/09/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-