MAUDE Adverse Event Report: COOK IRELAND LTD ECHOTIP ULTRA ENDOBRONCHIAL HD ULTRASOUND NEEDLE; FCG KIT, NEEDLE, BIOPSY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK IRELAND LTD ECHOTIP ULTRA ENDOBRONCHIAL HD ULTRASOUND NEEDLE; FCG KIT, NEEDLE, BIOPSY Back to Search Results
Catalog Number ECHO-HD-22-EBUS-P
Medical Device Problem Code Retraction Problem (1536)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 03/26/2017
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Pma/510(k) # k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.Investigation is pending.A follow up mdr will be submitted with the investigation conclusions.
 
Event or Problem Description
Mdr is being submitted based on the malfunction precedence of 'non-retraction of needle'.The ebus needle distal side become bent and almost broke during the procedure.The needle was replaced by other one.
 
Additional Manufacturer Narrative
Pma/510(k) # k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.The customers¿ complaint issue was reported as follows: ¿the ebus needle distal side become bent and almost broke during the procedure.The needle was replaced by other one.¿ additional information was received "regarding needle retraction-as much as i know and understood from the doctor, the needle was retracted to the sheath before removal." r&d provided a possible root cause: the root cause for this complaint is more than likely the physician tried to advance the needle into the tracheal rings causing the needle to bend.A caution note is in the product ifu stating the following ¿if excessive resistance is encountered on needle advancement, retract the needle into the sheath with the thumbscrew locked at 0cm mark, reposition the scope and attempt needle advancement from another angle.Failure to do so may result in device damage or malfunction¿.The device involved in this complaint was not available for return to cook ireland for evaluation.As the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.With the information provided a document based investigation was carried out.The customer complaint is considered to be confirmed based on customer testimony.Echo-hd-22-ebus-p devices of lot number c1192811 included the component irs0095 22ga ebus needle sub assembly of lot# ch1182719.Prior to distribution, all echo-hd-22-ebus-p devices are subjected to functional checks and visual inspection to ensure integrity of the product.These inspections and functional checks are outlined in internal procedures in place at cirl.A review of the manufacturing records for echo devices of lot# c1192811 did not reveal any discrepancies that could have contributed to this complaint issue.There is no evidence to suggest that this issue affects the entire lot # c1192811; upon review of complaints this failure mode has not occurred previously with this lot # c1192811.The notes section of the instructions for use ifu0060-2 that accompanies this device instructs the user to inspect the device prior to use for any damage: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".From the information provided to us the patient did not experience any adverse effects due to this occurrence complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
This follow up mdr is being submitted to include investigation conclusions.Additional information was received from the customer confirming that the needle could be fully retracted.However, from the image provided, it is very unlikely that this needle could be retracted into the sheath without perforating the sheath.Clarification has been sent out and still awaiting confirmation.Adopting a cautious approach an fda mdr report is still required based on the assumption that 'sheath perforation' occurred when needle was being retracted.The ebus needle distal side become bent and almost broke during the procedure.The needle was replaced by other one.
 
Additional Manufacturer Narrative
Pma/510(k) # k160229.Cook ireland ltd (manufacturer) is submitting this report on behalf of cook medical incorporated (cmi)(importer).Exemption number: e2016031.Information pertaining to section g.1 as follows: importer site contact and address: (b)(4) cook medical incorporated (cmi) 1025 acuff road p.O box 4195 bloomington indiana 47402-4195.Importer site establishment registration number: 3005580113.The customers¿ complaint issue was reported as follows: ¿the ebus needle distal side become bent and almost broke during the procedure.The needle was replaced by other one.¿ additional information was received "regarding needle retraction-as much as i know and understood from the doctor, the needle was retracted to the sheath before removal." r&d provided a possible root cause: "the root cause for this complaint is more than likely the physician tried to advance the needle into the tracheal rings causing the needle to bend.A bend like this could lead to retraction difficulties rather than sheath perforation / damage.The needle possibly could have been retracted with such a bend.It the customer stated there was no issue with retraction then we have to believe what they say right? they might have not thought it be difficult to retract whereas some other physician might disagree but i think it would be possible to retract the needle." engineering have provided additional feedback: "due to this device not being returned it is impossible to say if the user could retract it without perforating the sheath.Maybe it could be retracted without perforating the sheath but we can¿t say for certain.It could be possible." a caution note is in the product ifu stating the following ¿if excessive resistance is encountered on needle advancement, retract the needle into the sheath with the thumbscrew locked at 0cm mark, reposition the scope and attempt needle advancement from another angle.Failure to do so may result in device damage or malfunction¿.The device involved in this complaint was not available for return to cook ireland for evaluation.As the device was not returned for evaluation; the cause of this complaint could not be conclusively determined.With the information provided a document based investigation was carried out.Prior to distribution, all echo-hd-22-ebus-p devices are subjected to functional checks and visual inspection to ensure integrity of the product.These inspections and functional checks are outlined in internal procedures in place at cirl.There is no evidence to suggest that this issue affects the entire lot # c1192811; upon review of complaints this failure mode has not occurred previously with this lot # c1192811 the notes section of the instructions for use ifu0060-2 that accompanies this device instructs the user to inspect the device prior to use for any damage: "visually inspect with particular attention to kinks, bends and breaks.If an abnormality is detected that would prohibit proper working condition, do not use".From the information provided to us the patient did not experience any adverse effects due to this occurrence complaints of this nature will continue to be monitored for potential emerging trends.
 
Event or Problem Description
This follow up mdr is being submitted due to additional information received.Additional information was received from the customer confirming that there was no sheath perforation.Engineering have provided their feedback: "due to this device not being returned it is impossible to say if the user could retract it without perforating the sheath.Maybe it could be retracted without perforating the sheath but we can't say for certain.It could be possible." additional information was received from the customer on the (b)(6) 2017 contradicting the information the customer provided on the (b)(6) 2017 and the (b)(6) 2017.Adopting a cautious approach an fda mdr report is still required based on the assumption that 'sheath perforation' could still potentially occur due to the severe bend of the needle and that the customer's information is contradictory.The ebus needle distal side become bent and almost broke during the procedure.The needle was replaced by other one.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ECHOTIP ULTRA ENDOBRONCHIAL HD ULTRASOUND NEEDLE
Common Device Name
FCG KIT, NEEDLE, BIOPSY
Manufacturer (Section D)
COOK IRELAND LTD
o halloran road
limerick
MDR Report Key6536373
Report Number3001845648-2017-00158
Device Sequence Number1202104
Product Code FCG
UDI-Device Identifier00827002534087
UDI-Public(01)00827002534087(17)190120(10)C1192811
Combination Product (Y/N)N
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue NumberECHO-HD-22-EBUS-P
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Distributor Facility Aware Date05/02/2017
Event Location Hospital
Type of Report(Section G)FDA Requested
Initial Date Received by Manufacturer 04/03/2017
Supplement Date Received by ManufacturerNot provided
Not provided
Initial Report FDA Received Date05/02/2017
Supplement Report FDA Received Date05/30/2017
06/29/2017
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-