| Model Number |
Z-800W |
| Medical Device Problem Code |
Device Operates Differently Than Expected (2913)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
04/03/2017
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The manufacturer is still process of testing the infusion pump.
|
| |
|
Event or Problem Description
|
|
The user reported that the pump's flow rate was off.No patient injury or harm occured in this case.
|
| |
|
Additional Manufacturer Narrative
|
|
The user reported issue was not confirmed.The pump was found to have flow rate accuracy of -1.74%, which was within the +/-5% specification.The pump was found to have a battery outside of warranty, which was not related to the reported issue.Battery was replaced and the pump was tested to meet all specifications on 05/09/2017.
|
| |
|
Event or Problem Description
|
|
This is a follow-up for the initially filed mdr (3006575795-2017-00099).
|
| |
|