Model Number 383400
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
The unit was sent to a service center for repair.Upon triage, a technician found issues with the gearbox and sensor.The unit has been repaired, and recertified per procedure.The unit was restored to proper operation.Information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
The customer reports that the unit will not turn on, that it turns off after starting.
 
Manufacturer Narrative
An evaluation of the kangaroo joey pump was performed for the reported condition of ¿the unit turns off after starting.¿ the unit was triaged and the reported issue could not be confirmed at this time.A review of the device history record shows that this unit was manufactured in 2012 and was released meeting all manufacturing specifications.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
TUBES, GASTROINTESTINAL (AND ACCESSORIES)
Manufacturer (Section D)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN MEDICAL PRODUCTS
building 10- no 789 puxing roa
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6538307
MDR Text Key74255219
Report Number3006451981-2017-05207
Device Sequence Number1
Product Code LZH
UDI-Device Identifier10884521006645
UDI-Public10884521006645
Combination Product (y/n)N
Reporter Country CodeIT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number383400
Device Catalogue Number383400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received07/11/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/24/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-